METHODOLOGY
Reconstruction Techniques
Early approaches to 3D echocardiography were
based on the principle that a 3D data set could be
reconstructed from a series of 2D images. In this
method, serial 2D images are obtained using either
freehand scanning or a mechanically driven transducer
that sequentially recorded images at predefined
intervals.4-8 With freehand scanning, a series
of images is obtained by manually tilting the transducer
along a fixed plane, and a spatial locator
attached to the transducer translates the 3D spatial
location onto a Cartesian coordinate system.
This approach has several practical limitations,
including the relative bulk of the acoustic spatial
locators, which makes transducer manipulation
difficult, and the need for a clear and direct path
between the acoustic locators and the transmitter.
For electromagnetic spatial locator systems, an
additional problem is the potential for interference
of the electromagnetic field by ferromagnetic
material in close proximity to the transducer
(eg, material in hospital beds and medical equipment).
9
An alternative to freehand scanning is the use of a
mechanized transducer to obtain serial images at set
intervals in a parallel fashion or by pivoting around
a fixed axis in a rotational, fanlike manner. Because
the intervals and angles between the 2D images are
defined, a 3D coordinate system can be derived from
the 2D images in which the volume is more uniformly
sampled than with the freehand scanning
approach.
More recently, the use of a transesophageal or
transthoracic multiplane probe has emerged as a
readily available method to obtain rotational images
at defined interval angles around a fixed axis.10-15
Typically, images are collected over a 180-degree
rotation at set intervals. To minimize reconstruction
artifacts, sequential images are gated to both electrocardiography
(ECG) and respiration. Acquisition
of a complete data set typically takes 1 to 5 minutes,
depending on respiratory and heart rates and the
predefined spatial intervals. During cardiac surgery,
respiration can be suspended during acquisition to
minimize the effects of respiratory motion.
The quality of 3D reconstructions from 2D
images depends on a number of factors, including
the intrinsic quality of the ultrasound images, the
number (or density) of the 2D images used to
reconstruct the 3D image, the ability to limit
motion artifact, and adequate ECG and respiratory
gating. In general, the greater the number of
images obtained (ie, the smaller the space intervals
between images), the better the 3D reconstruction.
However, increasing the number of
images also lengthens the acquisition time, which
can potentially introduce motion artifact. Consequently,
the optimal number of images necessary
for 3D reconstruction depends on the cardiac
structure being examined and the resolution required.
For example, 4 to 6 serial images are
usually adequate for volume reconstructions of
the left ventricle (LV), whereas more images are
often needed to visualize more complex, rapidly
moving structures, such as mitral and aortic
valves.
Once the 2D images have been obtained, they
are processed offline with customized or commercially
available software. The cardiac structures
are manually or semiautomatically traced to the
3D spatial coordinates to reconstruct a 3D image.
No comments:
Post a Comment