Monday, August 30, 2010

CONCLUSION

CONCLUSION
Three-dimensional echocardiography is a safe, noninvasive
imaging modality that is complementary and
supplementary to 2D imaging and can be used to
assess cardiovascular function and anatomy in various
clinical settings. At present, available evidence suggests
that 3D echocardiography provides improved
accuracy and reproducibility over 2D methods for LV
volume and function calculation and the derivation of
mitral valve area in patients with mitral stenosis. Further
technological improvements and additional clinical
studies will broaden the list of appropriate applications
for this exciting new ultrasound modality.

FUTURE DIRECTIONS

FUTURE DIRECTIONS
Ongoing developments in 3D echocardiography
include technological innovations and expanding
clinical applications. Automated surface extraction
and quantification, single-heartbeat full-volume
acquisition, transesophageal RT3D imaging,
the ability to navigate within the 3D volume, and
stereoscopic visualization of 3D images are some
of the technological advances that can be expected
over the next several years. These will
further enhance the quality and clinical applications
of 3D echocardiography. In addition, standardized
and focused 3D protocols will be developed
and refined to optimize clinical application
of this technique.
Tagging and/or tracking the LV surface in real
time may provide new approaches to quantifying
myocardial mechanics, such as regional shape and
strain. This approach has great potential and will
complement and likely compare favorably with
the quantitative ability of cardiac MRI. The superior
temporal resolution of echocardiography
should offer unique advantages for this purpose.
In the future, combining the greater temporal
resolution of 3D echocardiography with the excellent
spatial resolution of MRI (or computed
tomography) may yield an imaging data set with

FUTURE DIRECTIONS
Ongoing developments in 3D echocardiography
include technological innovations and expanding
clinical applications. Automated surface extraction
and quantification, single-heartbeat full-volume
acquisition, transesophageal RT3D imaging,
the ability to navigate within the 3D volume, and
stereoscopic visualization of 3D images are some
of the technological advances that can be expected
over the next several years. These will
further enhance the quality and clinical applications
of 3D echocardiography. In addition, standardized
and focused 3D protocols will be developed
and refined to optimize clinical application
of this technique.
Tagging and/or tracking the LV surface in real
time may provide new approaches to quantifying
myocardial mechanics, such as regional shape and
strain. This approach has great potential and will
complement and likely compare favorably with
the quantitative ability of cardiac MRI. The superior
temporal resolution of echocardiography
should offer unique advantages for this purpose.
In the future, combining the greater temporal
resolution of 3D echocardiography with the excellent
spatial resolution of MRI (or computed
tomography) may yield an imaging data set with

Contrast Echocardiography

Contrast Echocardiography
The use of contrast with 3D echocardiography to
improve quantification of LV volumes offers several
advantages. The RT3D technique (single or
full volume) provides the most practical approach.
Triggering, although not essential, increases
the signal-to-noise ratio and thus is superior
to nontriggered imaging.38 Preliminary
clinical studies have shown promise with regard
to improved LV surface identification and volume
and ejection fraction measurement.129,130
Another evolving application of contrast 3D
echocardiography is in the evaluation of myocardial
perfusion. The ability to record the entire LV
and to quantify the full extent of hypoperfused
myocardium is a potential advantage of this approach.
131-133 However, the problem of microbubble
destruction, even with triggered imaging,
remains a challenge. This is especially true when
matrix array transducers are used, which results in
suboptimal myocardial opacification due to high
acoustic power. Further technological developments
should lead to improvements in all of these
areas and will contribute to more practical applications
of contrast 3D echocardiography.

Intraoperative Applications

Intraoperative Applications
The accuracy, feasibility, and value of 3D echocardiography
also have been demonstrated in the
intraoperative environment. Intraoperative 3D
echocardiography provides accurate and often
additional anatomic information compared with
2D transesophageal (TEE) imaging.123 In limited
studies examining 2D and 3D TEE intraoperative
evaluation of mitral valve prolapse anatomy, 3D
TEE evaluation provided complementary and additional
information compared with 2D TEE for
localization of prolapsed scallops (video clip
14).77,124 Intraoperative 3D TEE also has been
used to identify distortion and folding of the mitral
annulus as a cause of functional mitral stenosis or
worsening mitral regurgitation during beatingheart
surgery.125 Finally, intraoperative 3D TEE
has proven valuable in patients undergoing surgery
for congenital heart lesions. For example, the
superiority of intraoperative 3D TEE compared
with 2D has been demonstrated by its ability to
provide en face and oblique views of left atrioventricular
valve malformations in patients undergoing
reoperation for persistent regurgitant lesions
after previous repair of atrioventricular septal
defects.78,126
Intraoperative epicardial RT3D echocardiography
has been used to improve spatial orientation
and assess the extent of septal thickening, mitral
valve systolic anterior motion, and postsurgical LV
outflow tract patency in a patient with hypertrophic
cardiomyopathy undergoing septal myectomy,.
127 It also has been used to guide and monitor
off-pump atrial septal defect closure in a beatingheart
animal model.128 Finally, intraoperative epicardial
and postoperative transthoracic RT3D
echocardiography has been used to evaluate
changes in LV volume and function during cardiac
surgery in patients undergoing infarct exclusion
surgery for ischemic cardiomyopathy.69 In contrast
to 3D echocardiographic imaging, conventional
2D methods may not accurately quantify LV
volumes in patients with severe ischemic cardio

cardiomyopathy,
especially in the presence of significant
geometric changes due to LV aneurysm.

Congenital Heart Disease

Congenital Heart Disease
Clinical investigations examining the role of 3D echocardiography
in patients with congenital heart disease
have emphasized the unique perspective provided by
3D imaging and the versatility of the technique in

patients with simple defects or complex conditions
and in the postoperative state.115,116 Three-dimensional
echocardiography, using both reconstruction
methods and RT3D, has been used to detect several
forms of congenital heart disease. The ability to record
and analyze the entire cardiac structure and the ability
to display complex spatial relationships are potential
advantages of 3D imaging over 2D echocardiography.
In addition, the decreased examination time afforded
by RT3D echocardiography may reduce the need for
sedation in some children.116
In patients with atrial septal defects, 3D echocardiography
can record the size and shape of the defect. It
also can show the precise location of the defect and
the extent of residual surrounding tissue. In patients
with secundum atrial septal defects (Figure 11, video
clip 13), the extent of the retroaortic rim often determines
the feasibility of repair with percutaneous closure
devices. Three-dimensional echocardiography
also has been used after atrial septal defect closure to
evaluate the success of the procedure and identify the
origin of residual shunting.117 In patients with ventricular
septal defects, the ability to interrogate the entire
septum is frequently cited as an advantage of the 3D
technique.118,119 A novel application of 3D imaging in
patients with ventricular septal defects involves using
offline reconstruction to measure the shape and size of
the color flow jet, which allows for accurate measurement
of the magnitude of shunting in patients with
isolated ventricular septal defects.119
Various 3D echocardiographic techniques have
been used to evaluate RV and LV size and function in
patients with congenital heart disease. The approach
to the LV is similar to that described previously and
permits quantification of dimension, volume, mass,
and ejection fraction.120 Owing to the ellipsoidal shape
of the LV, the advantages of 3D over 2D echocardiographic
techniques are limited, because simple geometric
assumptions can be used to calculate LV volumes;
however, the RV’s asymmetrical shape
invalidates the simple geometric assumptions used for
LV volume calculations. In this case, the ability to
record and analyze the entire chamber rather than
relying on simplifying assumptions has proven
superior.48 In patients with congenital heart diseases
that involve RV pathology, 3D echocardiography correlates
well with MRI for the measurement of RV
volume.48,121,122
Three-dimensional echocardiography has been
successfully applied to the detection and assessment
of several anatomic defects. For example,
the circumferential extent and severity of discrete

subaortic membranes have been successfully visualized
with 3D echocardiography.119,120 With the
apical view, a unique en face image of the membrane
can be recorded, which permits analysis of
the effective orifice area and the dynamic nature
of the defect. Congenital malformations of the
mitral valve also have been assessed with 3D
echocardiography.80 The complex nature of these
defects can make a thorough anatomic evaluation
difficult. In such cases, the perspective provided
by 3D echocardiography can provide a complete
preoperative assessment of the extent and severity
of the valvular abnormality.

Valvular Heart Disease

Valvular Heart Disease
The recent widespread availability of RT3D echocardiography
obviates many of the practical limitations of
reconstructive 3D techniques and offers the potential
for greater clinical application for valvular heart disease
both in standard diagnostic evaluation and in
real-time guidance during surgical valve repair. This
technique is ideally suited for assessing valve function
given the nonplanar anatomy of the cardiac valves and
the associated anatomic and spatial alterations associated
with valvular heart disease.
Mitral Valve. The 3D echocardiography technique
has contributed significantly to our understanding of
mitral valve function and anatomy. The mitral valve is
particularly suited to 3D assessment because of the
complex interrelationships among the valve, chordae,
papillary muscles, and myocardial walls. This technique
can provide important insight into mitral valve
structure, demonstrating the saddle shape of the mitral
annulus, with high points located anteriorly and low
points oriented in a mediolateral direction (Figure 10,
video clip 8). This has helped clarify the appropriate

diagnostic imaging planes from which mitral valve
prolapse should be diagnosed, thereby avoiding falsepositive
interpretations.64,65
In addition, 3D echocardiography has provided important
mechanistic insights into functional and ischemic
mitral regurgitation resulting from derangements
of the normal spatial relationships of the mitral valve
leaflets to its chordal attachments, papillary muscles,
and the LV.13,66 Distortion of the normal spatial relationship
between the LV and mitral valve apparatus
results in papillary muscle displacement and tethering
of the mitral leaflets, leading to incomplete closure of
the leaflets and mitral regurgitation (video clip 9). The
3D echocardiography technique has identified
changes in annular shape occurring with functional
mitral regurgitation.67-69 These mechanistic and anatomic
insights based on 3D analysis have provided the
basis for the development of new approaches to
treating ischemic mitral regurgitation.70-74
Three-dimensional echocardiography has been used
to define and localize mitral leaflet lesions in mitral
valve prolapse, endocarditis, and congenital mitral
abnormalities.75-80 This application has been particularly
important in guiding surgical repair (video clips
10 and 11).81-83
The RT3D approach has also demonstrated efficacy
in quantifying mitral regurgitation by using 3D guidance
to directly measure the proximal flow convergence
region.84,85 It has provided insight into how
premitral orifice geometry affects the calculation of
mitral valve area in mitral stenosis.86 Calculation of
mitral valve area by 3D echocardiography has been
demonstrated to be accurate, reproducible, and less
variable than conventional 2D methods 

and thus has been recommended as the firstline
method.92 In addition, 3D echocardiography has
been used for guidance during percutaneous mitral
valvuloplasty.93,94
Aortic Valve. Three-dimensional echocardiography
has been applied for anatomic assessment of the aortic
valve and root morphology and to calculate the valve
area in aortic stenosis.95-99 The technique has been
used to delineate aortic flow patterns100,101 and has
demonstrated feasibility and accuracy in quantifying
aortic regurgitation.102,103 Other applications have included
the detection and localization of aortic valve
vegetations, assessment of congenital outflow obstruction
abnormalities, and demonstration of morphological
changes in the valve after balloon dilation.103-107
Tricuspid and Pulmonary Valves. Compared with
the aortic and mitral valves, the tricuspid and pulmonary
valves have been less widely studied with 3D
echocardiography. This technique has demonstrated
anatomic changes with rheumatic and degenerative
tricuspid valve disease.108-110 and has accurately reconstructed
congenital tricuspid valve abnormalities, such
as atrioventricular canal defects.111,112 For the pulmonary
valve, 3D assessment has been limited to descriptive
case reports defining anatomic abnormalities associated
with pulmonary valve stenosis and
endocarditis

CLINICAL APPLICATIONS

CLINICAL APPLICATIONS
Chamber Quantification
Left Ventricle. LV chamber and mass quantification
have been studied extensively using 3D echocardiography
(Tables 2 and 3).20-38 Initial 3D methods to
measure LV volumes used reconstruction techniques
that, although more accurate and reproducible than
2D methods, required long acquisition and postprocessing
times.20,21,30,35,39-41 Moreover, the accuracy of
the volume calculations was highly dependent on
image quality. The introduction of real-time imaging
systems that use matrix phased-array transducers with
more processing elements has significantly improved
image quality. In addition, LV quantification algorithms
that can interface with 3D data sets obtained with
matrix phased-array transducers are now widely available
and are increasingly robust.
The wide-angle acquisition mode is often used to
acquire the entire LV volume, from which a detailed
analysis of global and regional wall motion can be
done. Images may be displayed with either orthogonal

long-axis views or multiple short-axis views. Currently,
data analysis is performed offline on a personal computer
with dedicated 3D software. Data also can be
analyzed online with software intrinsic to the ultrasound
machine. Because a data set comprises the
entire LV volume, multiple slices from different orientations
can be obtained from base to apex to evaluate
wall motion. If image quality is limited, then acquisition
can be combined with infusion of contrast to
improve the delineation of the endocardial border. An
advantage of a 3D data set over 2D is the ability to
manipulate the plane to align the true long axis and
minor axis of the LV, hence avoiding foreshortening
and oblique imaging planes (Figure 7). Once the LV
axes are appropriately aligned, LV volumes can be
calculated with a centroid-based algorithm that typically
uses 2 or 3 planes (Figure 8), thereby shortening
processing time. In addition, the LV volumes can be
segmented, which allows for regional LV function
assessment (Figure 9, video clip 5).
LV volume assessment by RT3D has been demonstrated
to be rapid, accurate, reproducible, and superior
to conventional 2D methods.42 The superiority of
the RT3D approach has been demonstrated in various

clinical situations, but its use is limited in patients with
a poor acoustic window. An alternative method of
calculating ventricular volumes from an RT3D cardiac
volume data set uses the disc summation method. This
technique may be advantageous in patients with asymmetrical
ventricles.22,43 LV volume and mass obtained
by RT3D echocardiography compare favorably with
those obtained with cardiac magnetic resonance imaging
(MRI) or radionuclide volumes.39,42,44 In addition,
RT3D echocardiography has demonstrated efficacy
and accuracy in assessing LV volumes in remodeled
ventricles after myocardial infarction and in assessing
global LV dyssynchrony (video clips 6 and 7).32,45
Preliminary clinical studies on the use of RT3D in
stress echocardiography confirm the feasibility of this
technique and report sensitivity and specificity comparable
to 2D stress imaging.46,47 An advantage of
RT3D stress imaging is the decreased imaging time;
standard views can be obtained with only 1 or 2 image
acquisitions. In preliminary clinical studies, average
acquisition times decreased from 65 to 28 seconds
with RT3D imaging.46,47
Right Ventricle. Assessment of right ventricle (RV)
function by 2D echocardiography is limited because of

the asymmetrical, pyramidal shape of the RV, which
does not conform to simple geometric assumptions. In
theory, direct visualization of the entire chamber
should be possible with 3D techniques, thereby overcoming
the inherent limitations of tomographic methods.
To date, most studies that have applied 3D
echocardiographic techniques to the RV have involved
primarily rotational or freehand scanning methods
(Table 4); most of these series demonstrated improved
accuracy of RV function assessment.48-58 However,
these 3D data sets involved reconstruction from serial
2D images with the need for offline postprocessing,
thereby limiting their widespread clinical application.
The recent development and availability of RT3D
echocardiography has the potential to further improve
the ability to assess RV chamber size, volume, and
function.
Left Atrium. In a limited number of studies, left
atrial volume has been accurately quantified by 3D
echocardiography using both reconstructive and realtime
techniques. The 3D echocardiographic methods
correlate well with MRI59-62 and appear to have accuracy
comparable to 2D left atrial volume methods

Protocols

Protocols
A complete 3D echocardiographic study includes an
assessment of ventricular function, valvular morphology,
and hemodynamic status. Unlike 2D echocardiography,
in which standard views are described based on
the plane through which they pass, 3D echocardiography
is inherently volumetric. As such, it permits both
an external view of the heart and multiple internal
perspectives (through cropping).
Table 1 lists the components of a complete 3D
echocardiographic study. A general approach is to
describe cardiac structures using both the ultrasound
plane and the viewing perspective. Three
orthogonal planes are recommended: (1) the sagittal
plane, which corresponds to a vertical, longaxis
view of the heart; (2) the coronal plane,
which corresponds to a 4-chamber view; and (3)
the transverse plane, which corresponds to a
short-axis view (Figure 5). Each plane can be
viewed from 2 sides, which represent opposite
perspectives; for example, the transverse plane,

which represents the short-axis view, can be
visualized from the perspective of the apex or
base; the coronal plane can be viewed from above
or below; and the sagittal plane can be viewed
from the left or right. The choice of narrow-angle
or wide-angle imaging acquisition modes depends
on the cardiac structure to be examined. For
imaging of the ventricles, it is best to use a
wide-angle acquisition in the apical window (4-
chamber) so as to include the entire ventricle. For
smaller structures, such as the aortic valve, a
narrow-angle acquisition may be adequate.

As an alternative to a complete 3D study, 3D echocardiography
can be performed selectively as a complement
to a 2D study. Instead of a complete 3D
echocardiogram, a more focused 3D imaging study
may be appropriate in some cases. For example, in a
patient with mitral stenosis, the 3D portion of the
study may be limited to visualization and quantification
of the mitral orifice. Focused 3D imaging for LV
volume calculation, typically performed with an apical
4-chamber wide-angle acquisition, also can be used to
complement standard 2D imaging.
The ability to extract hemodynamic information
derived from 3D color Doppler ultrasonography is
currently being investigated. To capture and analyze
color flow imaging in 3 dimensions, the area of interest
should be obtained within the 3D data set, with the
angle of the ultrasound beam aligned as parallel as
possible to the direction of blood flow. Depth and
sector settings should be optimized for color Doppler
resolution. Extraneous flows can be cropped so that
only the area of interest is displayed. The color Doppler
flow patterns can be analyzed in multiple views to
provide a complete assessment of the color Doppler
data (Figure 6).

Technical Factors

Technical Factors
The technical aspects of acquiring a high-quality,
diagnostic 3D echocardiogram are similar to those
for 2D echocardiography. As with any new imaging
technique, a learning curve exists, and recognizing
and avoiding potential artifacts is critical.
Many of the artifacts are related to respiratory or
ECG gating and/or incorrect gain settings. The use
of optimal gain settings before acquisition is essential
for accurate diagnosis. Low gain settings
can artificially eliminate certain structures that
then will not be viewable during postprocessing.
Alternatively, using high gain settings can mask
structures and lead to significant misdiagnoses.
Therefore, overcompensating for the brightness
of the image using the time-gain compensations is
recommended, to allow the overall gain to be set
at midrange values. This maneuver will allow
maximum flexibility with postprocessing settings.
Most 3D echocardiographic systems use some
form of gating to obtain volumetric data. Gated
data sets are most challenging in patients with
arrhythmias or respiratory difficulties. The confounding
effects of the gating artifacts can be
minimized in different ways. For example, if the
gated system acquires sector slices in a sweeping
motion parallel to the reference image, then every
image viewed parallel to the reference image will
appear normal, whereas the gated artifacts will be
most noticeable when viewed from a plane orthogonal
to the reference image.
Segmentation is the process by which anatomic
features are extracted from the raw ultrasound

data. Segmentation can be accomplished using
low-level techniques, such as edge detection (in
2D) and surface detection (in 3D) based on local
features, such as the spatial gradient in ultrasound
intensity. More sophisticated techniques attempt
to extract entire boundaries or surfaces at once
based on local features and the anticipated shape
of the overall structure. Compression is a mathematical
technique that can be applied to the
original digital image file to reduce the amount of
data, thereby decreasing storage requirements and
improving retrieval rates. A single-volume data set
from a typical RT3D echocardiographic system
consists of 64 64 512 bytes (approximately 2
MB), or more than 50 MB for a 1-second loop, a
load that can overwhelm storage systems. Compression
of the digital data files can reduce this
load by about 3:1. The motion-JPEG algorithm
currently used by DICOM (Digital Imaging and
Communications in Medicine) and applied to individual
2D slices could be expected to achieve an
approximate 20:1 compression. More advanced
algorithms, such as wavelets (JPEG-2000), potentially
can yield better results. The use of compression
algorithms can decrease the size of data files,

optimizing storage efficiency without sacrificing
image quality.

Real-Time 3D Acquisition Methods

Real-Time 3D Acquisition Methods
Several studies have demonstrated that 3D reconstruction
from serial 2D images provides accurate
anatomic information suitable for quantitative
analysis.4-8,10-15 However, this methodology is subject
to technical limitations during image acquisition
and requires significant offline data processing.
The development of RT3D echocardiographic
systems circumvents many of the disadvantages of
reconstructive methods. RT3D echocardiography

uses a transducer with ultrasound elements arranged
in a grid fashion (Figure 2). The earliest
devices, developed by von Ramm and colleagues,
3,16,17 used a sparse-array matrix transducer
transmitting at a frequency of 2.5 or 3.5
MHz. These transducers consisted of 256 nonsimultaneous
firing elements and acquired a pyramidal
volume data set measuring 60° 60° within
a single heartbeat. However, the resolution and
image quality of this first-generation sparse-array transducer
were relatively poor and often inferior to standard
2D images; frame rates were low; and the pyramidal
volume had a narrow sector angle of 60°,
resulting in an inability to accommodate larger ventricles.
Moreover, the images obtained with this system
were not volume-rendered online; instead, they consisted
of computer-generated 2D cut planes derived
from the 3D volume data set. These features limited
clinical application of this pioneering technology.
Current RT3D systems use matrix-array transducer
technology with a greater number of imaging
elements, typically containing more than 3000
imaging elements, compared with the 256 in the
sparse-array transducer. These current matrix-array
transducers offer improved resolution and are
rapidly becoming the primary technique for 3D

data acquisition in clinical and research practice.
However, recent improvements in transducer
technology have resulted in (1) a smaller transducer
footprint, (2) improved side-lobe suppression,
(3) greater sensitivity and penetration, and
(4) harmonic capabilities that may be used for
both gray-scale and contrast imaging. In addition,
these matrix-array transducers display either online
3D volume-rendered images or 2 to 3 simultaneous
orthogonal 2D imaging planes (ie, biplane
or triplane imaging).
RT3D systems generally have 3 acquisition
modes: real time (narrow), zoom (magnified), and
wide angle. The real-time mode displays a pyramidal
data set of approximately 50° 30° (Figure
3A, video clip 1).18 The zoom mode displays a
smaller, magnified pyramidal data set of 30° 30°
at a higher resolution (Figure 3B). The wide-angle
mode provides a pyramidal data set of approximately
90° 90°, which allows inclusion of a
larger cardiac volume (Figure 3C, video clip 2).
This wide-angle mode requires ECG gating, because
the wide-angle data set is compiled by
merging 4 narrower pyramidal scans obtained
over 4 consecutive heartbeats. To minimize reconstruction
artifacts, data should be acquired during

suspended respiration if possible. Although wideangle
data sets provide a larger pyramidal scan,
this is at the cost of lower resolution, which is
decreased compared with the narrow-angle 3D
mode.
Once a 3D data set is acquired, it must be sliced or
“cropped” to visualize the cardiac structures within
the pyramid (Figure 4). Multiple cropping methods
are available, but a common method displays 2 or 3
imaging planes simultaneously (video clips 3a and
3b). Each of these imaging planes can be manipulated
separately to appropriately align the cardiac
structures. Another cropping method involves a
single-slice plane that can be manually adjusted to
expose and display the cardiac structures of interest

METHODOLOGY

METHODOLOGY
Reconstruction Techniques
Early approaches to 3D echocardiography were
based on the principle that a 3D data set could be
reconstructed from a series of 2D images. In this
method, serial 2D images are obtained using either
freehand scanning or a mechanically driven transducer
that sequentially recorded images at predefined
intervals.4-8 With freehand scanning, a series
of images is obtained by manually tilting the transducer
along a fixed plane, and a spatial locator

attached to the transducer translates the 3D spatial
location onto a Cartesian coordinate system.
This approach has several practical limitations,
including the relative bulk of the acoustic spatial
locators, which makes transducer manipulation
difficult, and the need for a clear and direct path
between the acoustic locators and the transmitter.
For electromagnetic spatial locator systems, an
additional problem is the potential for interference
of the electromagnetic field by ferromagnetic
material in close proximity to the transducer
(eg, material in hospital beds and medical equipment).
9
An alternative to freehand scanning is the use of a
mechanized transducer to obtain serial images at set
intervals in a parallel fashion or by pivoting around
a fixed axis in a rotational, fanlike manner. Because
the intervals and angles between the 2D images are
defined, a 3D coordinate system can be derived from
the 2D images in which the volume is more uniformly
sampled than with the freehand scanning
approach.

More recently, the use of a transesophageal or
transthoracic multiplane probe has emerged as a
readily available method to obtain rotational images
at defined interval angles around a fixed axis.10-15
Typically, images are collected over a 180-degree

rotation at set intervals. To minimize reconstruction
artifacts, sequential images are gated to both electrocardiography
(ECG) and respiration. Acquisition
of a complete data set typically takes 1 to 5 minutes,
depending on respiratory and heart rates and the

predefined spatial intervals. During cardiac surgery,
respiration can be suspended during acquisition to
minimize the effects of respiratory motion.
The quality of 3D reconstructions from 2D
images depends on a number of factors, including
the intrinsic quality of the ultrasound images, the
number (or density) of the 2D images used to
reconstruct the 3D image, the ability to limit
motion artifact, and adequate ECG and respiratory
gating. In general, the greater the number of
images obtained (ie, the smaller the space intervals
between images), the better the 3D reconstruction.
However, increasing the number of
images also lengthens the acquisition time, which
can potentially introduce motion artifact. Consequently,
the optimal number of images necessary
for 3D reconstruction depends on the cardiac
structure being examined and the resolution required.
For example, 4 to 6 serial images are
usually adequate for volume reconstructions of
the left ventricle (LV), whereas more images are
often needed to visualize more complex, rapidly
moving structures, such as mitral and aortic
valves.
Once the 2D images have been obtained, they
are processed offline with customized or commercially
available software. The cardiac structures

are manually or semiautomatically traced to the
3D spatial coordinates to reconstruct a 3D image.

BACKGROUND

BACKGROUND
Attempts to record and display ultrasound images in
3D format were first reported in the 1960s. One of
the earliest studies described the acquisition of a
series of parallel scans of a human orbit to reconstruct
3D anatomy.1 Despite the limited technology
of the day, these initial studies demonstrated that
complex anatomic structures were ideally displayed
using 3D techniques. Concerns about image quality
and the computational power needed for storage
and reconstruction greatly limited the early application
of this methodology.
More than a decade later, investigators began to
obtain 3D ultrasound images of the heart.2 Through
the careful tracking of a transducer, a sequence of
2-dimensional (2D) echocardiograms could be recorded,
aligned, and reconstructed into a 3D data
set. This methodology was limited by the need for
offline data processing to create and display the 3D
images. In the early 1990s, von Ramm and colleagues3
developed the first real-time 3D (RT3D)
echocardiographic scanner, capable of acquiring
volumetric data at frame rates sufficient to depict
cardiac motion. More recently, further improvements
in design and engineering have led to the
commercialization of RT3D echocardiography. This
methodology has evolved quickly, and different
versions of RT3D imaging are currently available on
several platforms.

3D Echocardiography: A Review of the Current Status and Future Directions

Ultrasound technology has improved markedly in
the past 10 to 15 years, prompting echocardiographers
to extend its use in studying cardiac
structure and function. New ultrasound equipment
and techniques offer superior image quality, greater
accuracy, and expanding capabilities. As a result,
more and improved imaging modalities are available
for evaluating cardiac anatomy, ventricular function,
blood flow velocity, and valvular diseases. Threedimensional
(3D) echocardiography offers the ability
to improve and expand the diagnostic capabilities
of cardiac ultrasound. However, as with any
emerging technology, the enthusiasm to embrace a
new technique must be tempered by a critical
appraisal of the evidence supporting its use. It is
essential to assess the limitations as well as the
unique capabilities it provides. Cardiac imaging
should be safe, accurate, versatile, comprehensive,
and cost-effective, while providing important clinical
information. Criteria for appropriate utilization
should be based on current evidence and updated as
new information becomes available.
To justify the use of a new 3D modality, its unique
contribution to clinical practice must be critically
analyzed. In this article we review the status of 3D
echocardiography, examine the evidence for its use
in various clinical situations, and propose guidelines
for appropriate application of this technique based
on available evidence.

The Future of 3D-Echo

The Future of 3D-Echo
With the rapid advances in digital image processing 3Dimaging
is probably just at the beginning of its evolution with
a number of innovations already approaching. The integration
of 3D-systems into conventional scanners and operator
friendly applications will reduce the time and effort required
to obtain 3D-images. Improvements in 2-dimensional
imaging and 3-dimensional reconstruction software will lead
to enhanced image quality. Novel ways of image representation
such as stereoscopy, holography or the generation of
physical 3D-models (Figure 6) could enhance our perception
of cardiac structures [17, 18].
Further advances in real-time 3-dimensional imaging and
the advent of real-time 3D-color Doppler will generate 3-dimensional
images directly on the screen during the investigation,
making the technique as simple as using Doppler or Mmode
features [19]. And, finally, other echocardiographic
modalities such as contrast echocardiography, tissue Doppler
imaging or intracardiac ultrasound could benefit from a
3-dimensional display format [20].

Some of these developments could soon become available
while others are still experimental. Despite present limitations
3-dimensional imaging could soon be considered as a
milestone in echocardiography along with M-mode, 2-dimensional
imaging or Doppler techniques.

Limitations

Limitations
However, despite the potential of 3D-echo to visualize cardiac
structures and perform volume computations this technique
has not gained wide spread acceptance to date. This
might be related to several factors: (1) 3D can only visualize
what is also seen on the two dimensional image, thus, an experienced
echocardiographer will obtain similar information
from a conventional examination without the need for costly
instrumentation and long post-processing times, (2) operator
experience with the reconstruction and interpretation of 3Dimages
is necessary, (3) 3D-image quality greatly depends on
the quality of the two-dimensional image and the ability to
obtain a motion and artifact free 3D-data set, (4) three-dimensional
imaging only creates a “virtual sense of depth” on a
flat (2-dimensional) screen. And finally, manual endocardial
contour tracing is still required to obtain 3D-volumes.
Some of these limitations will certainly be overcome with
newer techniques and growing experience with 3D-echo.

3D-Echocardiography as Teaching and Research Tool

3D-Echocardiography as Teaching and Research Tool
Spatial representation of cardiac structures greatly enhances
the understanding of cardiac function and pathology. Thus,
three-dimensional images could assist in the teaching
of echocardiography where a significant amount of spatial
understanding is required. An example of such an application
is a system which couples 3D-echo with a virtual reality heart
model [16]. The system allows standardized echocardiographic
views to be selected on the virtual heart and displayed
from the 3D-dataset to provide a correlation between
anatomy of the heart and echocardiographic image planes.

Potential Applications of 3D-Echocardiography

Potential Applications of 3D-Echocardiography
The potential applications of 3D-echo can be categorized
into 3 major areas: (1) Interpretation of morphology and
pathology, (2) Quantification of volumes and function,
(3) 3D-echocardiography as a teaching tool.
Interpretation of Morphology and Pathology
The clinical potential of 3D-echocardiography has been thoroughly
explored. Our own experience and that of others have
clearly demonstrated that the anatomy (Figure 1) and pathology
of the heart and the great vessels can often be displayed
[8, 9] in a more comprehensive format. Even fairly small
structures such as coronary arteries, a paravalvular leak or
small masses and vegetations can be visualized [7, 10].
Our findings also show that this technique can be applied
in numerous settings. For example, in valvular heart disease
(Figure 2), to determine the size of infectious vegetations, to
determine the mitral valve area in mitral stenosis (Figure 3),
for complex congenital malformations, or aortic dissection
(Figure 4). Furthermore, it has also been shown that jets can
be reconstructed from color Doppler information to assist in
the quantification of valvular lesions [11].
Quantification of Ventricular Volumes and Function
3D-echo has been applied to derive quantitative measurements
of volume, mass and dimensions of the left and right
ventricles and also other cardiac lesions, such as atrial and
ventricular septal defects [12]. While quantification of
ventricular volumes with two-dimensional imaging requires
geometric assumptions, measurement obtained with
3D-echo represents true volumes. Several studies have
shown 3D-echo to be superior to 2D-echocardiography for
both left and right ventricular volumes [13]. The processrequires acquisition of a 3-dimensional data set and manual
endocardial contour tracing. Several calculations including
volumes (throughout the cardiac cycle), global and regional
ejection fractions can be computed (Figure 5). The endocardial
surface of the ventricular cavity can be displayed from
multiple angles in a dynamic mode. Since the process of
manual endocardial border tracing is still time-consuming,
semi-automated contour detection algorithms are now being
developed. In addition, there is experimental evidence that
contrast opacification of the left ventricle could further
enhance the applicability of 3D-volume computation [14].
The advent of real time volumetric scanning will certainly
enhance the applicability of 3D-volume computation [15].

Principles of 3D-Echocardiography

Principles of 3D-Echocardiography
Data Acquisition
Three-dimensional echocardiography requires the collection
of a volumetric data set where each image (cut plane) is defined
with respect to its exact position in space [4]. Most systems
currently rely on sequential collection of image planes.
With this technique it is necessary to use ECG and respiratory
triggering or breath hold acquisition to account for motion
artifacts caused by respiration and to permit alignment of the
images in the time domain.
Newer developments use a transthoracic probe technology
with volumetric scanning capabilities, which allows simultaneous
acquisition of an entire 3D-data set. As a result,
data acquisition is less time consuming and less susceptible to
artifacts [5]. 3D-reconstructions have also been applied to
the color Doppler information allowing a three dimensional
representation of jets superimposed on the 3D-grayscale image.
Image acquisition can be performed from both a transthoracic
and a transoesophageal approach (TEE).
Transthoracic 3D-Echo
Three-dimensional transthoracic imaging can be performed
with mechanical steering devices, which are attached to
standard transducers. These devices steer the transducer motion
causing incremental changes in the scan plane either by
rotating, shifting or fanning the probe. In addition, various
locating systems (ie, acoustic or electromagnetic) have been
used effectively. The advantage of this technique is that freely
definable image planes can be chosen allowing for more flexibility.
Others have proposed a rapid (6 seconds) acquisition technique
that collects apical tomograms (within 6 sec) using an
internal continuously rotating transthoracic transducer [6].
Volumetric real-time echocardiography is a recently developed
technique based on the design of an ultrasound
transducer with a matrix array that instantaneously acquires
the image contained in a pyramidal volume. Volumetric realtime
echocardiography is a novel imaging concept, which
holds promise as a “break-through” technology for 3D-echo.
Employing a matrix array echo probe this technique allows
instant (real-time) acquisition of a complete 3-dimensionaldata
set without complex post-processing. Several studies
have already demonstrated the validity of real-time volumetric
echocardiography for the calculation of cardiac volumes
[5]. In addition, real-time volumetric echocardiography allows
the reconstruction of freely definable 2D-image planes
from a single volume set independently of the acquisition
window.
Transoesophageal 3D-Echo
First attempts to acquire a 3-dimensional data set from the
oesophagus were made with a specially designed probe
(echo-CT, lobster tail probe, Tom Tec). This probe was capable
of acquiring parallel data sets by passing a transducer
along the oesophagus [7]. Newer technologies, however, use
multiplane TEE probes that acquire sequential images at different
transducer rotation points (0–180o). 3D-echocardio-simply by mounting the steering device onto the TEE probe.
Data Post-Processing and Representation
Post-processing of the data for sequentially acquired images
is performed off-line using dedicated software. Varying
amounts of user interaction are required to define the region
of interest, view, cut plane, rendering algorithm, filter, magnification
and thresholds. The systems provide a variety of
3D-tools for advanced image processing. Multiplanar, 3Dreconstructions
(volume rendering) as well as wire frame
(surface rendering) display formats can be chosen and measurements
(distance, area, angle, volume) can be performed.
Recent advances in computing capabilities such as parallel
processing have greatly reduced the time necessary for data
manipulation. Three-dimensional reconstruction can now
be achieved within seconds and viewed from different angels
in a dynamic format. It is even possible to “electronically”
dissect the heart to visualize otherwise concealed structures.
Volumetric scanning allows instantaneous (real-time) display
of multiple views (multiplanar) using a split screen. In
addition, prototype systems have demonstrated the feasibility
of near real-time 3D-reconstruction, which permits almost
simultaneous display of 3D-images during the examination.

introduction

The interpretation of echocardiographic images requires a
complex mental integration of multiple image planes for
a true understanding of anatomic and pathologic structures.
The representation of images in a 3-dimensional format
more closely resembles reality and could therefore enhance
image interpretation. In addition, 3-dimensional imaging allows
direct calculation of volumes and is, thus, more accurate
than current models relying on geometric assumptions.
First attempts to incorporate multiple views to form a
3-dimensional image were made in the seventies. But, because
of technical limitations (eg lack of processing power,
relatively poor image quality, difficulties in image plane alignment)
this technique was limited to an experimental setting.
The advent of transoesophageal echocardiography together
with newer imaging probes and enhanced image
processing capabilities have now led to a remarkable progress
in the field of 3-dimensional imaging.
Numerous applications of three-dimensional echocardiography
(3D-echo) have been proposed. For example, improvements
in image interpretation with 3D-echo could be
of value in the decision making and planning of cardiac surgery,
and in the diagnosis of complex cardiac lesions [1]. In
addition, 3-dimensional imaging allows quantitative parameters
such as valve areas, the size of defects (atrial septal defect,
ventricular septal defect) or volumes to be obtained [2, 3].
With new developments that allow system integration of 3Dscanning,
rapid or even near real time 3D-reconstruction and
measurements, 3D-echo is now on the verge of becoming an
integral part of an echo examination.

Three-Dimensional Echocardiography – Principles and Promises

Three-Dimensional Echocardiography –
Principles and Promises
Three-dimensional echocardiography facilitates spatial recognition of intracardiac structures, potentially enhancing diagnostic
confidence of conventional echocardiography. In addition, 3D-echocardiography allows exact computation of cardiac
volumes and could serve as a teaching tool in cardiology. Over the last decade, significant developments in 3D-echocardiography
have been made. Refinements in instrumentation, data acquisition, post processing and computation speed together
with improvements in 2D-image quality could now allow three-dimensional echocardiography to play an important role in
clinical echocardiography. This review focuses on the methodology, current status, potential clinical applications and future
direction of 3D-echocardiography